EasyManuals Logo

Rigol DS1052D Service Guide

Rigol DS1052D
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
RIGOL Chapter 3 Performance Test
Service Guide for DS1000E, DS1000D Series
3-18
17) Repeat step 6).
18) Output a sine waveform with 20 MHz frequency and 1.2 Vpp amplitude from
Fluke 9500B.
19) Repeat step 8), 9) and 10).
20) Output a sine waveform with 50 MHz frequency and 1.2 Vpp amplitude from
Fluke 9500B.
21) Repeat step 12), 13) and 14).
22) Keep other settings of the oscilloscope unchanged and set the vertical scale
to 500 mV/div.
23) Output a sine waveform with 1 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
24) Repeat step 6).
25) Output a sine waveform with 20 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
26) Repeat step 8), 9) and 10).
27) Output a sine waveform with 50 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
28) Repeat step 12), 13) and 14).
29) Turn CH1 off. Test CH2 according to the above test steps and record the test
results.

Table of Contents

Other manuals for Rigol DS1052D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1052D and is the answer not in the manual?

Rigol DS1052D Specifications

General IconGeneral
BrandRigol
ModelDS1052D
CategoryTest Equipment
LanguageEnglish

Related product manuals