Chapter 5 Trigger RIGOL
MSO4000/DS4000 User’s Guide 5-49
position with a step of 1%. The range is from 5% to 95%.
6. Specify the trigger mode
Press Sweep to select “Auto”, “Normal” or “Single” (please refer to “Trigger
Mode”). The corresponding status light of the trigger mode currently selected in
the trigger control area (TRIGGER) at the front panel turns on.
7. Turn the noise rejection on or off
Press Setting to enter the trigger setting submenu. In this submenu, press
Noise Reject to turn the noise rejection on or off (please refer to “Noise
Rejection”).
Note: When the trigger source is D0 to D15, the noise rejection are invalid.