RIGOL  Chapter 5 To Trigger the Oscilloscope 
5-44  MSO7000/DS7000 User Guide 
Trigger Condition: 
Press  When,  and  rotate  the  multifunction  knob    to  select  the  desired  trigger 
condition,  then  press  down  the  knob  to  select  it.  You  can  also  press  When 
continuously or enable the touch screen to select it.   
⚫  Post: triggers on the specified position of the FlexRay bus. Press Post to select 
"TSS End", "FSS_BSS End", "FES End", or "DTS End". 
⚫  Frame: triggers on the frame of FlexRay bus. 
➢  Press  Frame to select  the frame  type. The types of  frames include null, 
Sync, Start, and All. 
➢  Press More → Define to select "ID" or "Cyc Count".   
When you select "ID", set the following parameters: ID Comp, ID Min, and 
ID Max. Press More → ID Comp to select the comparison conditions. The 
available choices include =,  ≠, >, <, ><, <>. Press More → ID Max or 
press More → ID Min, then rotate the multifunction knob    or use the 
pop-up numeric keypad to set the frame ID. 
When  you  select  "Cyc  Count",  set  the  following  parameters:  Cyc  Comp, 
Count  Min,  and  Count  Max.  Press  More  →  Cyc  Comp  to  select  the 
comparison conditions. The available choices include =,  ≠, >, <, ><, <>. 
Press More → Count Max or press More → Count Min, then rotate the 
multifunction knob    or use the pop-up numeric keypad to set the cycle 
count. 
⚫  Symbol:  triggers  on  the  CAS/MTS  (Collision  Avoidance  Symbol/Media  Access 
Test Symbol) and WUS (Wake Up Symbol) of FlexRay bus. 
➢  Press  Symbol  to  select  the  symbol  type.  The  symbol  type  includes 
CAS/MTS and WUS. 
➢  Press More → ID Comp to select the comparison conditions. The available 
choices include  =,  ≠,  >,  <,  ><,  <>.  Press  More  →  ID  Max  or press 
More → ID Min, then rotate the multifunction knob    or use the pop-up 
numeric keypad to set the frame ID. 
⚫  Error: triggers when an error occurs to the FlexRay bus. 
Press  Error  to  select  the  error  type.  It  includes  Head  CRC  Err,  Tail  CRC  Err, 
Decode Err, and Random Err. 
 
Note: As the occurrence possibility of specified FlaxRay frame is very low, it is 
recommended that you set the oscilloscope to "Normal" trigger mode when the 
trigger condition is set to "Frame", so as to prevent the instrument from triggering 
automatically while waiting for the specified frame. The same goes for "Error" trigger 
condition. In addition, when multiple FlexRay errors occur at the same time, you 
need to adjust the trigger holdoff so as to view the specific error. 
 
Trigger Mode: 
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch 
the current trigger mode. For details, refer to descriptions in "Trigger Mode".