RIGOL  Chapter 2 To Set the Vertical System 
2-10  MSO7000/DS7000 User Guide 
⚫  Bias Voltage 
The oscilloscope provides the bias voltage adjustment function for active probes. 
This function is used to adjust the signal under test that exceeds the input 
dynamic range of the probe amplifier to an appropriate range to ensure the 
integrity of the signal under test. Press Bias, rotate the multifunction knob   
or use the numeric keypad to set the DC bias voltage.   
 
⚫  Probe Information 
Press About to view information about the probe currently connected, such as 
the manufacturer, model, serial number, and the last calibration date. 
 
⚫  Probe Calibration 
Press Calibration to start zero calibration automatically for the probe.   
 
 
Amplitude Unit 
 
Select the amplitude display unit for the current channel. The available units are W, A, 
V, and U. When the unit is changed, the unit related to the relevant functions of the 
channel will change accordingly. 
 
Press 1 → More → Unit, press Unit continuously or rotate the multifunction knob 
  to select a unit. You can also enable the touch screen to select the desired unit 
with touch gestures. The default unit for the value is V. 
 
 
Channel Delay 
 
When using an oscilloscope for actual measurement, the transmission delay of the 
probe cable may bring relatively greater errors (zero offset). MSO7000/DS7000 
allows you to set a delay time for calibrating the zero offset of the corresponding 
channel. Zero offset is defined as the offset of the crossing point of the waveform 
and trigger level line relative to the trigger position, as shown in Figure 2-2.