RIGOL  Chapter 11 Protocol Decoding 
11-34  MSO7000/DS7000 User Guide 
⚫  When you select the analog channel (CH1-CH4), press Threshold, then 
rotate the multifunction knob    or use the pop-up numeric keypad to set 
the threshold of the source channel. When you modify the threshold of the 
channel, a dotted line displaying the current threshold level is displayed on 
the screen. The dotted line disappears in about 2 s after you stop modifying 
the threshold. 
 
4.  Signal setting 
Press Settings to enter the signal setting menu. 
⚫  Channel selection 
Press Channel continuously to select a channel ("A" or "B") that matches 
the actual FlexRay bus signal. 
 
⚫  Specify the signal rate 
Press Baud to open the baud rate list (2.5 Mbps, 5 Mbps, and 10 Mbps), 
and select a preset baud rate that matches the actual FlexRay bus signal. 
You can also press Baud continuously, then rotate the multifunction knob 
  at a small step value or use the pop-up numeric keypad to set the 
desired baud rate. 
 
⚫  Set the signal type 
Press Signal and rotate the multifunction knob    to select a signal type 
that matches the actual FlexRay bus signal. Press down the knob to select it. 
You can also press Signal continuously or enable the touch screen to select 
it. The available signal types include BP, BM, and RX/TX. 
 
⚫  Sample Position: 
Sample position is a point within a bit’s time. The oscilloscope samples the 
bit level at this point. The sample position is represented by the proportion 
of "the time from the start of the bit to the sample position" to the "bit time", 
as shown in the figure below. 
 
 
 
Figure 11-12 Sample Position 
 
Press More → Sample Position, then rotate the multifunction knob   
or use the pop-up numeric keypad to set the sample position. The settable 
range is from 10% to 90%.