ARBOTOM
®
- Manual 19
Connect the sensors using the sensor cables. Take into account that there are
different sockets for in- and output. The input socket is the one on the right side of
the sensor (), the output socket () on the left side. Always connect the output
socket () of a sensor with the input-socket of the next sensor.
Step 5: Connection the first sensor to the battery pack
Connect the input socket of the first sensor with the battery pack using the main cable
(5 or 10 meters).
Step 6: Control the correct connection of the sensors.
o Correct function:
When sensor chain is connected correctly, all sensors except the last one will show a
green LED when the battery pack is switched on. The LED of the last sensor will be
yellow.
Example: - - - - - - - - -
o Incorrect function:
In case of incorrect cable connection one or more intermediate sensors will show a
yellow LED. Please check the cable connection between the yellow sensor(s) and the
next one(s).
Example: - - - - - - - - -
In case of interrupted energy supply, an intermediate sensor will show a yellow LED.
Check the cable connection from this sensor to the next.
Example: - - - - X – X - X – X – X – X
Please always switch off the Battery Pack before making any changes of