RPR-4000
Operations manual
REV 8-21 29
Level Control CONTROL
When TRACK is set to Y, this parameter allows the operator to smoothly change the
output level of the gated amplifier. The parameter varies from 0 to 100, but does not
track exactly with output amplitude.
When TRACK is set to N, then the CONTROL field is not accessible to the operator.
The operator can directly control the RF gain and the bias level of the gated amplifier
using the controls RF and BIAS.
RF control voltage RF
This field is only accessible when TRACK is set to N. This control allows the operator to
change the RF gain in the low power sections of the RF amplifier.
BIAS control voltage BIAS
This field is only accessible when TRACK is set to N. This control allows the operator to
change the RF gain in the low power sections of the RF amplifier.
NOTE The controls RF and BIAS allow the user to make adjustments to the gain
and class of operation (A, AB, B etc.) of the gated amplifier. For guidance on using these
controls contact RITEC.
Burst Peak Voltage BURST PEAK VOLT
Displays the burst peak voltage. This readout is meant to be only an indication of the
voltage level, not a precise measurement.