To exit pass through mode, either send the command UP:<CR> via RS-232 communication
link or press the up arrow key on the local keypad
Upon exiting PASSTHROUGH mode, the local display will revert to the first menu, and the
last settings used during UPDATE mode will be reloaded into the pulser and receiver.
NOTE The maximum number of commands per second in Passthrough Mode is 50.
Sending multiple commands
A sequence of commands can be sent to the RPR-4000 by transferring a text file via the
serial port. The file must be a text file only.
NOTE When transferring the file, the lines must be sent with an appropriate delay in
order to stay below the limits of 20 commands per second for UPDATE
MODE and 50 commands per second for PASSTHROUGH MODE.