154
** Determined using measurement methods according to IEC 60601-2-24-1998.
General Technical Data
Maximum time for a Maintenance
message M24: OCCLUSION**
Self-filled cartridges:
j at medium basal rate 1.0 U/h: ≤2 h 10 min
j at minimum basal rate 0.02 U/h: ≤220 h
Pre-filled cartridges:
j at medium basal rate 1.0 U/h: ≤3h 10 min
j at minimum basal rate 0.02 U/h: ≤250 h
Maximum volume before Maintenance
message M24: OCCLUSION**
Self-filled cartridges: 1.0 U
Pre-filled cartridges: 2.0 U
Maximum pressure 350 kPa (3.5 bar)
Minimal opening pressure adapter vent 0.65 bar
Flow (delivery rate) Variable
Maximum quantity delivered at a single
fault condition
≤0.5 U
Cartridge
Self-filled 2.0 ml Accu-Chek Insight insulin pump cartridge
Pre-filled 1.6 ml NovoRapid
®
PumpCart
®
(please check the availability in your country at
www.novonordisk.com or ask your local pump support)