General Technical Data
Infusion sets Accu-Chek Insight infusion sets with a proprietary connector.
Remote control and data transfer Bluetooth wireless technology, ISM 2.4 GHz communication channel
Safety
The safety concept is based on a control system which consists of two microprocessors
and a supervisor microprocessor (supervising system).
The control system is based on a dual channel software architecture which performs all
safety relevant functions two times.
Whenever a defect or fault occurs in the control system, it is identified by the supervising
microprocessor and vice versa. The motor is immediately switched o and an Error E7:
Electronic error occurs.
The control system and supervising system signalize the Error E7 acoustically (beeps) and
tactile (vibrations). Additionally the control system shows the specific Error E7 number on
the pump display.
The motor also constitutes an important safety component, as the combination of the
control and supervising processors and brushless motor provides the best possible
reliability and accuracy in insulin delivery.
IPX8
Protected against the eects of temporary immersion in water under standardised
conditions (1.3 m for 60 minutes).

Table of Contents

Other manuals for Roche Accu-Check Insight

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Accu-Check Insight and is the answer not in the manual?

Roche Accu-Check Insight Specifications

General IconGeneral
BrandRoche
ModelAccu-Check Insight
CategoryMedical Equipment
LanguageEnglish

Related product manuals