156
Technical standards
Table 201: Electromagnetic emission
Guidance and manufacturer’s declaration – electromagnetic emissions
The Accu-Chek Insight insulin pump is intended for use in the electromagnetic environment specified below. The customer or the
user of the Accu-Chek Insight insulin pump should ensure that it is used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 2
The Accu-Chek Insight insulin pump must emit electromagnetic
energy in order to perform its intended function. Nearby electronic
equipment may be aected.
RF emissions
CISPR 11
Class B
The Accu-Chek Insight insulin pump is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable