13
Blood Glucose Tests
2
9
or
Blue
Test result is above
the target range.
Green
Test result is within
the target range.
Red
Test result is below
the target range.
Your test result appears with an arrow. This arrow shows if your test result falls
above, within, or below the target range.* The target range is represented by the
green region of the target range indicator. The arrow will flash if your test result
falls above or below this range.
* The target range is set by default at 70–180 mg/dL (3.9–10.0 mmol/L). To change
the target range, pair the meter to a mobile device, or connect it to a PC with
diabetes management software installed. Consult your healthcare professional
before changing the target range.
NOTE
When performing a blood glucose test: If the control bottle symbol and the
flashing L1 or L2 appear on the display with your blood glucose result, an error
has occurred.
• Do not act on the blood glucose result.
• Discard the test strip and repeat the blood glucose test with a new test strip.
Remove and discard the used test
strip by pulling the test strip out of the
meter or by pressing the test strip
ejector on the side of the meter.
Press here to eject test strip.
83715_09767959061A.indd 1383715_09767959061A.indd 13 6/29/2021 11:30:32 AM6/29/2021 11:30:32 AM