Roche/Hitachi cobas c 311 analyzer Appendix A - Instrument Alarms
Instrument Alarm List
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 G-6
Alarm Type Alarm No
Alarm Message Alarm Description Alarm Remedy
126 021 Host Communication
Error
No valid Record. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
126 022 Host Communication
Error
First Record is not a Header Record. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
126 023 Host Communication
Error
Undefined Record. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
126 024 Host Communication
Error
Data other than specified. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
126 025 Host Communication
Error
Data other than specified. a. Check the host computer. Is it ON?
b. Verify Host Communication on Start Conditions.
c. Check cable connections between the analyzer and host computer.
d. Check the host computer transmit condition.
e. Ensure that the host and the analyzer are utilizing the same communication configuration.
f. Resume operation; if alarm recurs, call Technical Support.
340 1 Host Receiving Text Error A field pause of L record is insufficient. Please check L record.
340 2 Host Receiving Text Error There is no record end of L record. Please check whether '<CR>' (0x0d) is in the terminus of L record.
340 3 Host Receiving Text Error Termination Code of L record is not effective
value.
Please check Termination Code of L record.
340 4 Host Receiving Text Error There is no record end of P record. Please check whether '<CR>' (0x0d) is in the terminus of P record.
340 7 Host Receiving Text Error A field pause of O record is insufficient. Please check O record.
340 8 Host Receiving Text Error Sequence Number of O record is not effective
value.
Please check Sequence Number of O record.
340 18 Host Receiving Text Error There is no record end of O record. Please check whether '<CR>' (0x0d) is in the terminus of O record.
340 19 Host Receiving Text Error Report Type is not effective value. Please check Report Type.
340 20 Host Receiving Text Error A field pause of Q record is insufficient. Please check Q record.
340 21 Host Receiving Text Error Sequence Number of Q record is not effective
value.
Please check Sequence Number of Q record.
340 22 Host Receiving Text Error Starting Range ID Number is not effective
value.
Please check Starting Range ID Number.
340 23 Host Receiving Text Error There is no record end of Q record. Please check whether '<CR>' (0x0d) is in the terminus of Q record.