cobas c 111
Host Interface Manual
Version
2
.1
44 / 93
Field Name Reference Format
Comments (Data Content) O|9.4.
Instrument Specimen
ID:
Sample ID / Sequence
^Rack ID / Carrier ID
^Position on
Rack/Carrier
^Tray ID / Location
ID
^Rack/Carrier type
^Tube/Cont. type
09.4.04 CM Identification for samples analyzer host.
cobas c 111 instrument receives data: Field is ignored (09.4.03 is
read)
cobas c 111 instrument transmits data:
The sample identification known by the instrument
Format
<Sample ID>^^<Position ID>
Element Max Length Format
Sample ID 23 ST
Position ID 1 NM
Elements & Description
Sample ID
Samples:
Instrument’s sample-order identification, e.g. either the barcode, a
sequence # (returned unchanged or a new ID if is entered on instrument)
Controls:
‘1000 + Control Code (ACN/BCN)’ is used.
e.g.
PNU: Control No. = 300, is sent as “1300”
See also action code field 09.4.12
Position ID
Values 1 to 8. Counting from the sample area left to right positions first
sample is 1
Position on the sample area if the sample is still on-board, otherwise
empty.