EasyManuals Logo

Roche Cobas C111 User Manual

Roche Cobas C111
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
cobas c 111
Host Interface Manual
Version
2
.1
44 / 93
Field Name Reference Format
Comments (Data Content) O|9.4.
Instrument Specimen
ID:
Sample ID / Sequence
^Rack ID / Carrier ID
^Position on
Rack/Carrier
^Tray ID / Location
ID
^Rack/Carrier type
^Tube/Cont. type
09.4.04 CM Identification for samples analyzer host.
cobas c 111 instrument receives data: Field is ignored (09.4.03 is
read)
cobas c 111 instrument transmits data:
The sample identification known by the instrument
Format
<Sample ID>^^<Position ID>
Element Max Length Format
Sample ID 23 ST
Position ID 1 NM
Elements & Description
Sample ID
Samples:
Instrument’s sample-order identification, e.g. either the barcode, a
sequence # (returned unchanged or a new ID if is entered on instrument)
Controls:
‘1000 + Control Code (ACN/BCN)’ is used.
e.g.
PNU: Control No. = 300, is sent as “1300”
See also action code field 09.4.12
Position ID
Values 1 to 8. Counting from the sample area left to right positions first
sample is 1
Position on the sample area if the sample is still on-board, otherwise
empty.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cobas C111 and is the answer not in the manual?

Roche Cobas C111 Specifications

General IconGeneral
BrandRoche
ModelCobas C111
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals