.1
45 / 93
Field Name Reference Format
Comments (Data Content) O|9.4.
Universal Test ID:
^^^Test ID
^Treatment
Type (and value)
^Pre-Treatment
Type (and value)
^Result Evaluation
Type (and value)
09.4.05 CM
repeat
Repeat Field (example: ^^^2^1\^^^100\^^^744^1)
Specifies Test Orders.
Format ^^^<HostCode>^<Dilution>
The first three elements for this format are empty. Currently Universal Test
IDs are not supported.
This format is used repeatedly with delimiters when ordering several tests.
Maximum of 60 tests can be specified.
Element Max Length Format
Test ID = Host Code Number
Ω
5 NM
Treatment 3 ST
Element Description
Test ID, Profile ID (Host Application Code):
This is a key field for orders. When transmitting results, this field is left
empty, i.e. the relevant Test ID is part of the Result Record.
The analyzer identifies the test order with 3-digit numbers.
Range : 1-999
NOTE: the Host Application Code is translated from and into the cobas c
111 instrument ACN space using a configurable table.
Treatment:
General: Specifies treatment to be done on instrument: (A)utodilution,
(D)ilution and factor, (C)oncentration and factor, etc., e.g. A, D100, etc.
(Roche specific field).
Note: concentration of samples is not supported on c111.
cobas c 111 instrument receives data: Any Value but NULL dilution /
concentration is done according to the value entered in the application
protocol
If NULL - pipetting and testing is done using the standard analysis
parameters.
Pre-Treatment & Result Evaluation:
Not supported on cobas c 111 instrument
Priority 09.4.06 ST Indicates the priority order of the Patient Samples. Ignored for Control
Samples. Values are one of the following:
Field Value Description
R Indicates routine analysis of the patient samples
Ψ
S Indicates Stat analysis of the patient samples
Ω
The record should be sent with this field left blank when the host has no Test Selections. See
example on page 36.
Ψ
The record should be sent with R when the host has no Test Selections. See example on page 36.