Field Name Reference Format
Comments (Data Content) R|10.1.
Data or Measurement
Value
10.1.04 ST When “result status" = ‘X’ (i.e. No Value) no data is set in this field.
Element Description
Quantitative results:
The value corresponds to the measured concentration:
e.g., 1.97, 123.4, etc.
In case a result is above / below a limit (measurement range), the
characters ‘<’ or ‘>’ are added in front of the result (limit) e.g., <0.05,
>10.0, etc.
Qualitative results: ARE NOT AVAILABLE ON cobas c 111 instrument
Units 10.1.05 CM Example: mg/dL
Format;
<UnitString>
Indicates units of the analytical data.
Reference Ranges 10.1.06 NM Format: LowerLimit/UpperLimit/OtherRange
Relates to field 10.1.07:
L or H: Below/Above reference range (Reference Range),
< or >: Low/High scale of assay (test range)
For QC samples: target value/ranges for
Lower limit/Upper limit see example1 page 50
The system submits the values entered in the application protocol, if no
value is specified the corresponding field is left empty.
Result Abnormal
Flags
10.1.07 ST Field Value Description
L : Less than reference range
H : Higher than reference range see example page 50
< : Below absolute low scale of assay (test range)
> : Above absolute high scale of assay (test range) ex. page 50
N : Normal
A : Abnormal Technical problems
Roche instruments provide more instrument specific flags than those
defined above by ASTM. The detailed flag information is transmitted to the
Host using an additional Comment Record.
Nature of Abnormality
Testing
10.1.08 Not used / ignored by cobas c 111 instrument