Proceed as follows to use the R&S FSH-Z28 or
R&S FSH-Z29 calibration standards:
Select FSH CALKIT in the CALKIT menu with the
rotary knob or the cursor keys.
To perform phase correction for additional cables and
adapters proceed as follows:
Select the menu item OFFSET LENGTH… in the
CALKIT menu using the rotary knob or the cursor
keys.
Terminate the selection with the ENTER key or the
MEAS MODE softkey.
Using the cursor keys or the rotary knob, change the
value for the additional electrical length of a cable or
adapter or enter a value with the numeric keys and
terminate with the ENTER key.
The additional electrical length is now taken into
account for phase measurements and in the Smith
chart.
Renaming the USER calibration standard:
The setting for the USER calibration standard can be assigned a user-defined name. The name of the
USER calibration standard which has been entered is then shown in the R&S FSH status display
(STATUS key) so that, for example, the setting can be documented when the measurement is
documented.
Select Rename USER in the CALKIT menu with the
rotary knob or the cursor keys.
Terminate the selection with the ENTER key or the
MEAS MODE softkey.
The R&S FSH opens the entry window for the name of
the USER standard.
Enter a name with the numeric keys.
Terminate the entry with the ENTER key.
When the CALKIT menu is called, the name that has
been entered is displayed in the USER menu item, e.g.
calkit-1 (USER).