R&S FSH Trace Settings
1309.6275.12 4.21 E-2
Operating sequence:
Press the TRACE key.
Press the DETECTOR softkey.
The submenu for selecting the detector opens.
If automatic operation has been selected, the AUTO
DETECTOR menu item is marked with X and the
R&S FSH displays the detector that is set to match the
trace mode that is set.
To switch automatic operation on or off:
Using the cursor keys or rotary knob, select AUTO
DETECTOR from the menu.
Using the ENTER key or the DETECTOR softkey,
switch automatic operation on or off.
When automatic operation is switched on, the
R&S FSH also sets the detector that matches the trace
mode that is set.
To set the detector manually:
Using the cursor keys or the rotary knob, select the detector you want and confirm by pressing the
ENTER key or the DETECTOR softkey.
The R&S FSH indicates the detector that has been selected in the top right-hand corner of the screen
(Detect: Auto Peak in Fig. above). If AUTO DETECTOR is switched on and a detector that does not
accommodate automatic operation is set, the R&S FSH will switch automatic operation off.
Trace-Selection
The R&S FSH can display two active traces TRACE 1 and TRACE 2 on the screen simultaneously.
Each active trace is allocated a trace memory (MEMORY 1 or MEMORY 2). To establish to which
measurement curve the control inputs are to refer, you must select a trace. All inputs relating to Trace
mode, Detector, and copying of the trace into the Trace memory relate to the selected trace. In the
default setting, TRACE 1 is selected for inputs.
Operating sequence:
Press the TRACE key.
Press the SELECT TRACE softkey.
Trace 2 is set instead of Trace 1. The selected trace is
shown in the upper area of the screen (Trace 1 / Trace
2)
Now apply the desired settings for Trace 2 (TRACE
MODE / DETECTOR / TRACE > MEMORY).
To change the settings of Trace 1, press the
SELECT TRACE softkey again.