Using Limit Lines R&S FSH
1309.6275.12 4.34 E-2
Exiting the list of limit lines:
Press the EXIT softkey to close the list of limit lines.
Switching off a limit line:
Press LINES.
Press softkey UPPER LIMIT / LOWER LIMIT.
In the menu select DESELECT LIMIT using the rotary knob or the cursor keys and confirm with
ENTER.
After a limit line has been switched on, the R&S FSH
returns to the menu and the selected line is displayed
in the diagram. The name and type of the limit line are
also marked (UPPER LIMIT for an upper limit line and
LOWER LIMIT for a lower line).
All active limit lines can be switched off together with the LIMITS ON/OFF softkey.
Measurements with limit lines
During a measurement, the R&S FSH checks the trace after each frequency sweep for upper and lower
limit violations. If all measured values are within specified limits, PASS is displayed at the top in the
center of the diagram. FAIL is indicated even if only a single measured value (= pixel of the trace)
exceeds a limit value.
The automatic limit check can be switched off via softkey LIMITS ON/OFF. A limit violation can also be
indicated by an acoustic signal.
Beep:
Press the OPTIONS softkey.
Select AUDIO BEEP with the rotary knob or the
cursor keys and confirm the selection with the
OPTIONS softkey or with the ENTER key.
If BEEP has been selected, the R&S FSH outputs a
beep each time a limit is exceeded.