Measuring the occupied bandwidth R&S FSH
1309.6275.12 4.44 E-2
Press the LEVEL ADJUST softkey.
The R&S FSH starts the measurement of the optimal
reference level using a resolution bandwidth of 3 MHz,
a video bandwidth of 3 MHz and the peak detector.
While the measurement is in progress, the R&S FSH
outputs the messageAdjusting level for
measurement, please wait... ".
The optimal reference level is then set.
Setting the channel bandwidth
The channel bandwidth determines the span, resolution bandwidth and sweep time the R&S FSH uses
for measuring the occupied bandwidth.
Press the CHANNEL BW softkey.
An entry box showing the current channel bandwidth
setting opens.
Using the numeric keys, enter a new channel
bandwidth and terminate the entry with the
appropriate unit; or
Change the channel bandwidth with the rotary
knob or the cursor keys and terminate the entry
with the ENTER key or the CHANNEL BW softkey.
The R&S FSH automatically adapts the span to the
channel bandwidth that has been entered (span = 5 x
channel bandwidth) to ensure that no incorrect
measurements of occupied bandwidth are made. The
minimum channel bandwidth that can be set is 2 kHz.
If you attempt to enter a smaller channel bandwidth,
the R&S FSH will automatically set 2 kHz.
Entering the power percent to determine the occupied bandwidth
Press the % POWER BW softkey.
The R&S FSH will open a field for entering the power
percent relative to the total span power which defines
the occupied bandwidth (percent of total power). The
R&S FSH displays the value currently set.
Using the rotary knob or the cursor keys, change
the percent value, or enter a value using the
numeric keys, and confirm the entry with the
ENTER key or the % POWER BW softkey.
The R&S FSH will now display the occupied bandwidth
of the specified percent of the total power.