R&S FSH Measuring forward and reflected power
1309.6275.12 4.77 E-2
1 Directional Power Sensor R&S FSH-Z44
2 Source
3 Load
4 Power sensor socket
When measuring high powers, strictly observe the following instructions to prevent damage to the
sensor or hazard to persons:
CAUTION
Danger of skin burns or damage to the instrument
Never exceed the permissible continuous power.
See diagram on the rear of the sensor for the permissible continuous
power.
Turn off the RF power to connect the sensor.
Screw the RF connectors tightly.
Operating sequence:
Press the MODE key.
Press the POWER METER softkey.
The power meter submenu opens.
The R&S FSH opens the screen for power measurements. If no power sensor is connected, it does not
display any measured value and outputs POWER SENSOR (UNKNOWN) in the status field. If a
power sensor is connected, the R&S FSH sets up a connection via its interface, displays first the
message POWER SENSOR (DETECTING) and then the message POWER SENSOR (BOOTING) in
the status field and, after a few seconds, displays the type of the sensor connected (R&S FSH-Z44) as
well as the measured power.