R&S FSH Setting the Amplitude Parameters
1309.6275.12 4.11 E-2
Press the AMPT key.
Press the TRANSDUCER softkey.
The R&S FSH opens a submenu to select the primary
and secondary transducer factors.
Using the rotary knob or the cursor keys, select the
menu item SELECT PRIMARY TRANSDUCER and
confirm with the ENTER key.
The list of factors available as primary transducer
factor opens.
Using the rotary knob or the cursor keys, select the
transducer factor and confirm the selection with the
SELECT softkey.
The selected transducer factor will then be displayed in
addition the entry field of the reference level.
To select another transducer factor, use the menu
entry SELECT SECONDARY TRANSDUCER and
proceed as described above.
To switch off a selected transducer factor, proceed as
follows:
Press the AMPT key.
Press the TRANSDUCER softkey.
The R&S FSH opens a sub-menu, in which you can
select primary and secondary transducer factors.
In the submenu select PRIMARY TRANSDUCER or
SECONDARY TRANSDUCER with the rotary knob
and confirm the selection with the ENTER key.
The transducer factor will then no longer be included in
the level display.