Preparing for Use
R&S
®
FPL1000
22Getting Started 1323.1602.02 ─ 07
To connect the AC power
1. Plug the AC power cable into the AC power connector on the rear panel of the
instrument. Only use the AC power cable delivered with the R&S FPL1000.
2. Plug the AC power cable into a power outlet with ground contact.
The required ratings are listed next to the AC power connector and in the data
sheet.
For details on the connector, refer to Chapter 5.2.2, "AC Power Supply Connec-
tion and Main Power Switch", on page 43.
4.5.2 Connecting an Optional DC Power Supply (R&S FPL1-
B30)
The R&S FPL1000 can also be equipped with an optional DC power supply con-
nector (R&S FPL1-B30). If installed, the R&S FPL1000 can be operated by a DC
voltage of +12 V to +24 V.
If you use an external power supply unit to supply safety extra-low DC voltage
(SELV) to the instrument, be sure to meet the requirements for reinforced/double
insulation in accordance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No.
1010.1) or DIN/EN/IEC 60950 (UL 1950, CSA C22.2 No. 950). Provide current
limitation in accordance with DIN EN 61010-1 Appendix F2.1.
Also see "Connecting to power" on page 8.
DC connection
Connect the DC power connector on the rear panel of the R&S FPL1000 to
the DC power source.
Connecting to Power