Operating the Instrument
R&S
®
FPL1000
75Getting Started 1323.1602.02 ─ 07
GAT The frequency sweep is controlled via the "TRIGGER INPUT" connector.
TDF The specified transducer factor is activated.
75 Ω The input impedance of the instrument is set to 75 Ω.
FRQ A frequency offset 0 Hz is set.
CWSource:
<level>
Internal generator in operation as independant CW source with specified level
(requires option R&S FPL1-B9).
<"NOR" |
"APX">
/Trk.Gen
Internal generator in operation as tracking generator (requires option
R&S FPL1-B9).
NOR: the measurements are normalized with the results of the internal genera-
tor calibration
APX (approximation): the measurements are normalized with the results of the
internal generator calibration; however, the measurement settings have been
changed since calibration
If neither label is displayed, no calibration has been performed yet or normali-
zation is not active.
For details, see the "Internal Generator" section in the R&S FPL1000 User
Manual.
Changing the Channel Setup Name
The channel setups are labeled with their default name. If that name already
exists, a sequential number is added. You can change the name of the channel
setup by double-tapping the name in the channel bar and entering a new name.
7.1.2 Window Title Bar
Each channel in the R&S FPL1000 display can contain several windows. Each
window can display either a graph or a table as a result of the channel measure-
ment. The window title bar indicates which type of evaluation is displayed.
Trace Information in Window Title Bar
Information on the displayed traces is indicated in the window title bar.
2
1 3 4 5
(1) Trace color
Color of trace display in diagram
(2) Trace no.
Trace number (1 to 6)
Understanding the Display Information - Spectrum Mode