EasyManua.ls Logo

Rohde & Schwarz RTE1104 - Page 1307

Rohde & Schwarz RTE1104
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1307User Manual 1326.1032.02 ─ 20
MEASurement<m>:ONViolation:PRINt <Print>
Prints a screenshot including the measurement results to the printer defined using
SYSTem:COMMunicate:PRINter:SELect<1..2> for the specified event.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<Print> NOACtion | SUCCess | VIOLation
See Chapter 18.4.5, "Event parameter", on page 1075
*RST: NOACtion
Usage: Asynchronous command
MEASurement<m>:ONViolation:WFMSave <SaveWfm>
Saves the waveform data.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<SaveWfm> NOACtion | SUCCess | VIOLation
See Chapter 18.4.5, "Event parameter", on page 1075
*RST: NOACtion
Usage: Asynchronous command
MEASurement<m>:ONViolation:REPort <Report>
Creates and saves a report of the current settings and results.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<Report> NOACtion | SUCCess | VIOLation
See Chapter 18.4.5, "Event parameter", on page 1075
*RST: NOACtion
Usage: Asynchronous command
Automatic measurements

Table of Contents

Related product manuals