Data and file management
R&S
®
RTE
498User Manual 1326.1032.02 ─ 20
To export waveform data of a running acquisition
1. Select the waveforms to be saved and the scope as described in "To save a wave-
form or a part of a waveform to a file" on page 497, step 1 to 5.
2. If you want to save only a section of each waveform, set the "Scope".
3. Enable "Data logging".
4. Enter the number of acquisitions to be acquired and saved in "Acq count".
5. Check the filename under "Save to file" and change it, if needed. Usually, autonam-
ing is used.
6. Tap "Start Export" to start the acquisition and to save the acquired waveform data
to the specified file.
To load waveform data as a reference waveform
To reload waveform data from a previous measurement, the waveform must have been
stored as a reference waveform in a BIN file before.
The procedure is described in "To load a reference waveform" on page 291