Remote control commands
R&S
®
RTE
1675User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..4
Parameters:
<SignalRevType> DISabled | DIFFerential | SINGle
*RST: DISabled
BUS<m>:HBTO:THReshold <ThresholdUpper>
Sets an user-defined threshold value.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdUpper> Range: 0 to 2
Increment: 1E-3
*RST: 0.4
Default unit: V
BUS<m>:HBTO:POLarity <Polarity>
Selects the polarity of the data signal.
Suffix:
<m>
.
1..4
Parameters:
<Polarity> NORMal | INVert
*RST: NORMal
BUS<m>:HBTO:MODE <Mode>
Selects the direction of the full-duplex signal you want to look at.
Suffix:
<m>
.
1..4
Parameters:
<Mode> MASTer | SLAVe | AUTO
*RST: AUTO
BUS<m>:HBTO:SYMRate <SymbolRate>
Defines the rate of ternary symbols.
Suffix:
<m>
.
1..4
Protocols