Quick Installation Guide
00825-0100-4725, Rev BA
July 2003
Rosemount 8732/8700 Series
STEP 4: INSTALLATION (WAFER FLOWTUBE)
Gaskets
The flowtube requires a gasket at each of its connections to adjacent
devices or piping. The gasket material selected must be compatible with the
process fluid and operating conditions. Metallic or spiral-wound gaskets
can damage the liner. Gaskets are required on each side of the
grounding ring. All other applications (including flowtubes with lining
protectors or a grounding electrode) require only one gasket on each
end connection.
Size
Code
Line Size
Polyurethane, Linatex, and Neoprene Liners
PN 10 PN 16 PN 25 PN 40
(Newton-
meter)
(Newton)
(Newton-
meter)
(Newton)
(Newton-
meter)
(Newton)
(Newton-
meter)
(Newton)
010 1 inch
(25 mm)
20 7040 20 7040
015 1.5 inch
(40 mm)
30 10700 30 10700
020 2 inch
(50 mm)
40 13400 40 13400
030 3 inch
(80 mm)
30 11100 30 11100
040 4 inch
(100 mm)
40 11700 50 13200
060 6 inch
(150mm)
60 16400 90 19200
080 8 inch
(200 mm)
90 23400 60 13100 90 19400 110 22800
100 10 inch
(250 mm)
70 18600 80 18800 130 25400 170 29900
120 12 inch
(300 mm)
80 21300 110 25500 130 25800 180 31900
140 14 inch
(350 mm)
110 29100 150 33000 210 38200 280 45400
160 16 inch
(400 mm)
150 33700 190 37400 280 45400 410 62000
180 18 inch
(450 mm)
130 28700 230 45600 220 36800 280 42700
200 20 inch
(500 mm)
150 34100 260 45900 300 48800 350 49400
240 24 inch
(600 mm)
200 39200 380 62400 390 60100 560 74400
4725_revBA_qig.fm Page 13 Thursday, July 3, 2003 12:47 PM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rosemount 8732 Series and is the answer not in the manual?

Rosemount 8732 Series Specifications

General IconGeneral
CategoryMeasuring Instruments
DiagnosticsYes
Measurement TypeVolumetric Flow
Output4-20 mA, HART
Enclosure RatingNEMA 4X, IP66
Communication ProtocolHART, FOUNDATION Fieldbus, Modbus
Product TypeTransmitter
EnclosureStainless Steel
Process ConnectionFlanged
MaterialStainless Steel
CertificationsATEX, IECEx, CSA, FM
Power Supply11 to 42.4 VDC
Flow Range0.04 to 12 m/s (0.13 to 39 ft/s)