EasyManuals Logo

Sakura Cyto-Tek 2500 User Manual

Sakura Cyto-Tek 2500
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
Page 11
9. Open the lid by pressing the Lid Open button
(Figure 2H). Remove the guard bowl unit from the
instrument (Figure 2I).
Figure 2H
Figure 2I
10. Remove a foam cushion (Figure 2J).
Figure 2J
11. Install the lid gasket, see Section 4.1.3. (Figure 2K).
12. Replace the guard bowl into the instrument
(Figure 2K).
Figure 2K
Lid open push button
Guard bowl unit
Foam cushion
Lid gasket
Guard bowl unit