EasyManua.ls Logo

Sam 271 - Precautions; Flammable Gases and Anesthetics

Default Icon
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Product Support 1-888-202-9831
DO NOT administer treatment if the applicator is not connected to a sam
Ultrasound Coupling Patch.
Applicators and sam Ultrasound Coupling Patches are not sterile. DO NOT apply
this device over an open wound or inflamed skin.
DO NOT use the sam Ultrasound Coupling Patch if the sam Ultrasound Media
is dried out. Indications of a dried out Patch include: the cup is not full of gel,
there is dry residue or film in the cup, or there is any cut, break, or opening in
the Patch or seals.
DO NOT apply directly over a bone that is near the skin surface.
3.3. PRECAUTIONS
Precaution should be taken when using the device:
Over an area of the spinal cord following a laminectomy, i.e. when major covering
tissues have been removed
On patients with hemorrhagic diatheses
Over areas where metal prosthesis or other metallic implants are embedded
in tissue which may form a reflective surface to the ultrasound energy causing
unintended irradiation of tissue and excessive heating
Over an acute infection or sepsis
On patients with peripheral artery disease
Over a deep vein thrombosis
Over an anesthetized area or in conjunction with a condition that causes
impairment of sensation, such as caused by chemotherapy
When using the sam Ultrasound Coupling Patch, ensure the top and bottom
seals have been completely removed before attaching the applicator to the skin
3.4. INFLAMMABLE GASES AND ANESTHETICS
Warning: Explosion hazard if used in the presence of flammable anesthetics,
open flame, or oxygen-rich environment

Related product manuals