EasyManua.ls Logo

S&A CWFL-6000 - Read only parameters (state)

S&A CWFL-6000
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CWFL-6000 Series
15
4. Read only parameterstate
Read-only
Register
address
Description
Bit definition
0x0800
Relay output
status
MSByte
Bit7(MSb)
Retain
Bit6
Retain
Bit5
Retain
Bit4
Retain
Bit3
Retain
Bit2
Alarm output relay
0Off 1On
Bit1
Conductivity lower limit relay
0Off 1On
Bit0(LSb)
High temp. heating rod
0Off 1On
LSByte
Bit7(MSb)
High temp. water pump relay
0Off 1On
Bit6
High temp. refrigeration relay
0Off 1On
Bit5
Compressor crankcase
heating
0Off 1On
Bit4
Low temp. heating rod
0Off 1On
Bit3
Low temp. refrigerant
solenoid valve
0Off 1On
Bit2
Condensing fan
0Off 1On
Bit1
Compressor relay
0Off 1On
Bit0(LSb)
Low temp. water pump
0Off 1On
0x0801
Digital input
switch status
MSByte
Bit7(MSb)
Retain
Bit6
Retain
Bit5
Retain
Bit4
Retain
Bit3
Retain
Bit2
Retain
Bit1
Retain
Bit0(LSb)
Retain
LSByte
Bit7(MSb)
Retain
Bit6
Retain
Bit5
External input
0No input 1Input
Bit4
Refrigerant low pressure
0No input 1Input
Bit3
Refrigerant high pressure
0No input 1Input
Bit2
Spare 2
0No input 1Input
Bit1
Spare 1
0No input 1Input
Bit0(LSb)
Water level
0No input 1Input

Related product manuals