- 34 -
Tem
p
erature distribution
Temperature of the cycle in each area (SV=-70
air temperature)
Unit:
Maximum Minimum
Middle o
f
c
y
cle
Differential
Left front -67.6 -74.7 -71.2 ±3.6
Left bac
k
-69.3 -78.3 -73.8 ±4.5
Ri
g
ht front -66.0 -73.3 -69.7 ±3.7
Ri
g
ht bac
k
-68.4 -77.2 -72.8 ±4.4
Cente
r
-68.0
-75.5 -71.8 ±3.8
Left front -66.7 -70.3 -68.5 ±1.8
Left bac
k
-69.5 -75.0 -72.3 ±2.8
Ri
g
ht front -66.1 -69.6 -67.9 ±1.8
Ri
g
ht bac
k
-69.5 -74.3 -71.9 ±2.4
Cente
r
-68.8
-72.6 -70.7 ±1.9
Left front -62.4 -64.8 -63.6 ±1.2
Left bac
k
-64.3 -67.5 -65.9 ±1.6
Ri
g
ht front -61.7 -63.9 -62.8 ±1.1
Ri
g
ht bac
k
-62.1 -65.1 -63.6 ±1.5
Cente
r
-63.8 -66.1 -65.0 ±1.2
-69.0 -72.7 -70.9 ±1.9
-66.1 -67.7 -66.9 ±0.8
---68.7 -
Maximum Minimum
Middle o
f
c
y
cle
Differential
Left front -68.6 -77.4 -73.0 ±4.4
Left bac
k
-66.9 -73.9 -70.4 ±3.5
Ri
g
ht front -65.6 -72.9 -69.3 ±3.7
Ri
g
ht bac
k
-68.0 -76.5 -72.3 ±4.3
Cente
r
-67.5
-75.0 -71.3 ±3.8
Left front -66.0 -69.8 -67.9 ±1.9
Left bac
k
-69.0 -74.6 -71.8 ±2.8
Ri
g
ht front -65.5 -69.0 -67.3 ±1.8
Ri
g
ht bac
k
-68.9 -74.0 -71.5 ±2.6
Cente
r
-68.2
-72.3 -70.3 ±2.1
Left front -61.3 -63.5 -62.4 ±1.1
Left bac
k
-63.4 -66.4 -64.9 ±1.5
Ri
g
ht front -60.5 -62.7 -61.6 ±1.1
Ri
g
ht bac
k
-61.1 -64.4 -62.8 ±1.7
Cente
r
-62.8 -65.0 -63.9 ±1.1
-68.2 -72.0 -70.1 ±1.9
-65.2 -67.0 -66.1 ±0.9
---68.0 -
Amount of power consumption
Amount of power consumption when driving at cyc
l
(
SV=-70䉝䠅
UnitkWh/da
y
Ambient temperature 30
Ambient temp. 30
Center of Lower shel
f
230V/50Hz
8.57
230V/50Hz
9.96
230V/50Hz
Ambient temp. 20
Bottom are
a
Center of Upper shel
f
Average
Note:This data does not represent a guarantee of product performance.