- 3 -
Item Specifications
Cascade sensor temperature is higher than 䠉12䉝, L side compressor
turns off.
Cascade sensor temperature is lower than 䠉34䉝, L side compressor
turns on.
Compressor protection
Overload relay
When filter sensor temperature is higher than +60䉝, H side compressor
turns off. (Substitution for pressure switch)
Compressor delay time
Range: 3~15 minutes (Unit: 1minute)
It is not settable delay time individually between H side compressor and L
side compressor.
Performance specifications
Item Specifications
Maximum cooling performance
-86䉝 at the center of the chamber 䠄AT䠖30䉝䚸No load䠅
Temperature control range
䠉50䉝 to 䠉86䉝 䠄AT䠖30䉝䚸No load䠅
Power source 230V, 50Hz 240V, 50Hz
Rated power consumption 650W 700W
Noise level 47dB <A> (background noise; 20dB)
Maximum pressure 2030 kPa