EasyManua.ls Logo

Sartorius 663002N - Maintenance Schedule; Greasing Components (1-Channel Models)

Sartorius 663002N
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Picus
®
2 Operating Instructions 47
Cleaning and Maintenance
t When a total of ten measurements have been recorded: Convert the
recorded masses (mi) into volume (Vi) by multiplying them with a
correction factor Z. The following applies: Vi=mi · Z (for Z scores, see
Chapter14.11, page58).
t Calculate the average volume: (V): V = (Vi)/10.
t To evaluate conformity: calculate the systematic error “eS” of the
measurement. Use the following formula:
in µL: eS = VVS
or in %: eS = 100 (V–VS)/VS
VS = selected test volume
t Calculate the random error of the measurement as the standard
deviation. Do this using the formula. n = number of measurements (10).
t Compare the systematic error and the random error with the values of
the performance specification from your own laboratory.
9.5 Maintenance Schedule
Interval Component Activity
6 to 24 months, depend-
ing on the operating con-
ditions
Device Contact Sartorius Service.
Regularly | depending on
the operating conditions
Tip cone Grease the components
(see 9.6, 47).
9.6 Greasing Components (1-channel models)
Requirement
The lower assembly is dismantled.
Procedure
t Wet the brush with the grease supplied.
For models with 3µL | 10µL | 120µL
t Apply a thin layer of grease to the piston (1).
For models with 300µL | 1,000µL
t Apply a thin layer of grease around the seal (1).
For models with 5,000µL | 10,000µL
t Apply a thin layer of grease inside the tip cone and around the seal (1).
1
1
1

Table of Contents