16
Information on Weighing Electro statically Charged Samples
lf you are using a balance with a 0.1 mg readability and electro stati-
cally charged glass or plastic Containers with a diameter larger than
the weighing pan, you should utilize our antistatic pan
(Ø 130 mm - see "Accessories") instead of the Standard weighing
pan.
Leveling the Balance Using the Level Indicator (depends on type
of model)
At the point of use, level the balance using the leveling feet
(8)
so that
the air bubble is centered within the circle of the level indicator
(10)
.
To raise the balance, lower the leveling feet (turn clockwise).
To lower the balance, raise the feet (turn counter-clockwise).