Using Verified Balances as Legal Measuring Instruments in the EU*
You must calibrate the balance at the place of installation before using it
as a legal measuring instrument (see the section entitled “Adjustment/
Calibration” starting on page 22).
This balance is not allowed to be used for weighing goods intended for direct
sale to the public. The type-approval certificate for verification applies only
to non-automatic weighing instruments. For balances of accuracy class k,
a thermometer and barometer are recommended for monitoring ambient
conditions. The temperature range indicated on the verification ID label must not
be exceeded during operation.
The balance must warm up for at least 24 hours after initial connection
to AC power.