Printout/Record for Adjustment or Calibration and Linearization Functions
A printout or record is generated at the end of the following functions:
all calibration, adjustment and linearization procedures
calibration test
The printout can have the following lines:
--------------------
MC1 - Sartorius : Balance family and manufacturer
Model BP211D : Balance model
S/N 040500048 : Serial number of the balance
ID : Space for entering the workstation or operator ID
--------------------
Date : 21-Mar-97 : Current date
Start: 10:05:30 : Time at which the application started
Cal. : Extern : Calibration mode (in this case,
“external calibration”)
Set. : 200.00000 g : Calibration weight value
(only for “external calibration”)
End : 10:05:45 : End of application
Name : : Space for signature of the operator responsible
--------------------
Additional information for other calibration modes:
Cal. : Intern : Internal calibration
Stat.: Complete : Status comment for calibration or linearization
functions
Cal. : Test : Calibration test
Diff.:- 0.00004 g : Data measured during calibration test