Tare Parameter
You can define when the balance will perform the taring operation:
Code**
At any time 1 5 1
Not until the readout is stable 1 5 2*
Auto Zero Function
When this zero-tracking function is activated, any slight changes off the zero
readout are automatically tared.
Code
Auto Zero on 1 6 1*
Auto Zero off 1 6 2
Adjustment/Calibration and Linearization Functions Using CAL
Internal Linearization applicable only to the BP 211 D, 301 S (-0CE)
Code
External calibration 1 9 1*/**
Internal calibration for balances with a built-in
automatic calibration function 1 9 3*
Calibration test for balances with a built-in
automatic calibration function 1 9 4
Internal linearization with automatic calibration 1 9 6
Adjustment/calibration functions blocked 1 9 7
** = factory setting; depends on the balance model in some cases
** = s ett ing no t ap plicable to verifi ed b alances o f accu rac y class K
Important Note Concerning Verified Balances Approved for Use as Legal
Measuring Instruments in the EU:
For verified balances of accuracy class K, the menu access switch is adjusted
to lock the “External Linearization” function after verification, and a control seal is
affixed to the balance to lock the menu access switch.