Balance Operating Parameters
Adapting the Balance to Ambient Conditions
The balance can be adapted to the prevailing ambient conditions at the place
of installation.
Code
Very stable conditions 1 1 1*
Stable conditions 1 1 2*
Unstable conditions 1 1 3
Very unstable conditions 1 1 4
Standard Weighing Mode ā Manual Filling Mode
You can optimally adapt your balance to meet either of these requirements.
In the manual filling mode, the display compensates for fluctuations of the load on
the balance, giving you especially fast and stable readouts.
Code
Standard weighing mode 1 2 1*
Manual filling mode 1 2 2
Stability Range
The stability symbol will remain displayed in the case
of a weight variation +/ā Code
0.25 digit 1 3 1
0.5 digit 1 3 2*
1digit
1
)133
2digits
1
) 1 3 4*
4digits
1
)135
8digits
1
)
2
)136
* = factory setting; depends on the balance model in some cases
1
) = not applicable to the BP 8-0CE and BP 2100-0CE
2
) = not applicable to the verified balances approved for use as legal measuring
instruments in the EU
31