Model BP 221S BP 161P BP 121S BP 61S
Weighing range structure SuperRange PolyRange SuperRange SuperRange
Weighing capacity g 220 30/60/ 120 61
110/160
Readability g 0.0001 0.0001/ 0.0001 0.0001
0.0002/
0.0005/
0.001
Tare range (by subtraction) g – 220 –160 –120 – 61
Reproducibility (standard deviation)* g ≤±0.0001 ≤±0.0001/ ≤±0.0001 ≤±0.0001
0.0002/
0.0005/
0.001
Linearity g ≤±0.0002 ≤±0.0002/ ≤±0.0002 ≤±0.0002
0.0002/
0.0005/
0.001
Response time (average) s ≤2
Allowable ambient operating temperature °C +5 ... +40
Operating temperature range °C +10 ... +30
Sensitivity drift within +10 °C ... +30 °C /°C ≤±1·10
–6
Pan size mm Ø 80
Weighing chamber height mm 225
Dimensions (WxDxH) mm 204 x297x332
Net weight, approx. kg 5.4
External calibration weight
(of at least accuracy class...) g 200 (E2) 100 (E2) 100 (E2) 50 (E2)
Power requirements V~ Via AC adapter: STNG 6, 230 VAC or 115 VAC,
–20% ... +15% (IP 20 protection)
Frequency Hz 48–63
Power consumption (average) VA 16 maximum; 8 average
Hours of operation with fully charged
YRB 05 Z external battery pack, approx.: h 32
Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
Analytical draft shield xxxx
Dust cover xxxx
AC adapter, varies acc. to country xxxx
Built-in calibration weight xxxx
Level indicator xxxx
Hanger for below-balance weighing xxxx
* = standard deviation of the reproducibility acc. to DIN 8120, Part 3
69