40
The difference between the weight value and the true mass is displayed, with +/- sign.
External calibration Calibration record is printed, if adjustment was not performed and the process was
Nom. + 10000 g stopped by pressing (
Diff. + 1 g
) Activate calibration/adjustment (press the ( key to cancel).
The calibration weight is displayed at the conclusion of calibration
-------------------- A GMP-compliant printout is generated
24.10.2002 10:15
Typ CIS3
Vers. 1.0010.10.2
BVers. 01-26-01
--------------------
External calibration
Nom. + 10000 g
Diff. + 1 g
External adjustment
Diff. + 0 g
--------------------
24.10.2002 10:15
Name:
--------------------