79
The display shows Altitud for 2 seconds.
The elevation at the place of installation is displayed in meters above sea level.
In the example shown here, the elevation setting for “Germany (Zone D)” is displayed.
) Confirm the displayed value or press ( to cancel the calibration procedure.
The display shows
Latitud for 2 seconds.
The geographical latitude of the place of installation is shown in degrees north
or degrees south.
In the example shown here, the latitude setting for “Germany (Zone D)” is displayed.
) Confirm the displayed value or press ( to cancel the calibration procedure.
You are prompted to place the required weight on the platform (e.g., 5.0 kg).
The subsequent steps for completing the calibration/adjustment are described in
the chapter entitled “Operation,” under “Calibration and Adjustment.”
In place of
Altitud and Latitud, Gravity is displayed for 2 seconds if the gravitational
acceleration was entered rather than the elevation and geographical latitude (menu:
“Calibration/Adjustment” - “Geographical Data” - “Input Parameters”).
Extern. cal. factory
-
def. w
Extern. cal. factory
-
def. w
Extern. cal. factory
-
def. w