EasyManua.ls Logo

Sartorius Combics 3 CAIS3 - External Calibration with User-Defined Weight; Configuration for User-Defined Weight; Performing External Calibration with User Weight

Sartorius Combics 3 CAIS3
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
External Calibration/Adjustment with a User-Defined Weight
Configuration
Select the “
CAL key function:Ext. cal./adj.; user-def.
wt." menu item (external adjustment using a user-defined weight).
Note:
The menu items “
Ext. cal./adj.; factory-def. wt." (external
adjustment with a factory-defined, standard weight), “
Ext. cal./adj.;
user-defined wt." (external adjustment with a user-defined weight) and
Key blocked" can also be accessed without activating the Service mode.
t Use the “Q" or “q" soft key to select this menu item and confirm with “l."
y The menu item is marked as set by a circle “o."
t Enter the target value of the adjustment weight in the
Calibration/adjustment" menu under “Parameter for
external weight" in the “Cal/adj. wt" input field.
t Press e to turn off the device.
t Press e turn the device back on.
y The Sartorius logo is displayed briefly, after which the device is in normal weighing
mode.
t Press ( to unload and zero the scale.
t Press J to start the external adjustment.
y The display “C.EXT.U" appears briefly.
Note
If the display of geographical data (elevation and latitude or gravity) is activated, this
data is displayed and confirmed each with a press of the J key (you can cancel the
calibration/adjustment process using the ( key). See also “Entering Geographical Data
for Use in Legal Metrology" in this chapter.
Procedure
t Press J to confirm the displayed value or press ( to cancel the adjustment.
y The target value of the required adjustment weight (6.000 kg in the example) is displayed
as a negative value.
t Place the required adjustment weight on the platform.
If the calibration/adjustment sequence is set to “Cal. then auto adjust”, refer to the note
under “External Calibration/Adjustment with Factory-Defined Weight (Default Weight).”
y After a brief pause, the difference since the last span adjustment is displayed
(calibration).
t If you only want to perform a calibration, press the ( key to cancel the
calibration/adjustment procedure.
t Press J to adjust the scale.
y At the conclusion of the calibration procedure, the calibration weight is displayed as
a positive value.
36
Configuring Weighing Platforms
C.EXT.U
C.EXT.USE.

Table of Contents

Related product manuals