EasyManua.ls Logo

Sartorius Combics 3 CAIS3 - User-Defined Linearization Completion; Completing Linearization Steps

Sartorius Combics 3 CAIS3
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
43
Configuring Weighing Platforms
t Press J to apply linearization weight 1 or press ( to cancel the linearization function.
y After the linearization weight 1 has been saved you will be prompted to place the second
linearization weight on the weighing pan.
t Repeat the procedure for all required linearization weights.
y After the last linearization weight has been saved you will be prompted to remove any load
from the weighing pan.
t Remove all linearization weights from the weighing platform.
y The zero point is applied automatically after a brief time. The indicator automatically
switches to weighing operation.
When linearization is complete, the GMP-compliant printout shown on the left is generated.
t Slide the menu access switch to the right (= “closed" position).
--------------------
14.01.2010 13:00
Type CAW3P1-6DC-LCE
Ser.no. 12345678
Vers.no. 1.02.101110
BVers. 01-63-02
--------------------
Linearization
Wt.1 + 1.500 kg
Wt.2 + 3.000 kg
Wt.3 + 4.000 kg
Wt.4 + 6.000 kg
completed
--------------------
14.01.2010 13:02