Connecting Electronic Peripheral Devices
§ Make sure to unplug the balance from AC power
before you connect or disconnect a peripheral device
(printer or PC) to or from the interface port.
Warmup Time
To deliver exact results, the balance must warm up
as listed below after initial connection to AC power
or after a relatively long power outage.
Model CPA2P…, CPA26P…: at least 4 hours
All other precision and analytical models:
at least 30 minutes
Only after this time will the balance have reached
the required operating temperature.
Using Verified Balances in Legal Metrology:
$ Allow the equipment to warm up for at least
24 hours after initial connection to AC power.
Antitheft Locking Device on Balances with
a Weighing Capacity of up to 10 kg
G To secure the balance at the place of installation,
fasten a chain or a lock to the lug located on the
rear panel of the balance (order no.: LC1).
13