114  Operating Instructions Cubis
®
 MCA
  Technical Data
15.7.5  Models MCA623S | MCA623P | MCA323S | MCA5202S
MCA623S  MCA623P MCA323S MCA5202S
Unit Value Value Value Value
Scale interval (d) mg 1 1 | 2 | 5 1 10
Maximum capacity (Max) g 620 150 | 300 | 
620
320 5200
Repeatability at 5% load
Standard deviation of the load values, 
tolerance
mg 0.7 1 0.7 6
Standard deviation of the load values, 
typical value
mg 0.4 0.4 0.4 2
Repeatability at approx. maximum capacity
Standard deviation of the load values, 
tolerance
mg 0.7 1 0.7 6
Standard deviation of the load values, 
typical value
mg 0.5 0.5 0.5 2
Linearity deviation
Tolerance mg 2 5 2 10
Typical value mg 0.6 1.5 0.6 5
Deviation when load is off-center, 
positions according to OIML R76
Test weight g 200 200 200 2000
Tolerance mg 2 4 2 10
Typical value mg 1 3 1 5
Sensitivity drift between +10°C and +30°C ppK 2 2 2 2
Tare maximum capacity: Less than 100% 
ofmaximum capacity
Accuracy class according 
toDirective2014/31/EU
II II II I
Verification scale interval (e) according 
toDirective 2014/31/EU
mg 10 10 10 100
Minimum load (Min) according 
toDirective2014/31/EU
mg 20 20 20 1000
Minimum initial weighing according to USP 
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 820 820 820 8200
Typical minimum initial weighing mg 820 820 820 8200
Typical stabilization time s 0.8 0.8 0.8 0.8
Typical measurement time s 1 1 1 1