EasyManuals Logo

Sartorius Cubis MCA225S Series User Manual

Sartorius Cubis MCA225S Series
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
Operating Instructions Cubis
®
MCA 75
System Settings
7.7.14 Parameters in the “Connections / SBI Protocol” Menu
Parameter Setting Values Explanation
Format Value without header The data output only exports the measured value without ID code.
Value (with header)* The data output exports the measured value with ID codes.
Date & time, value
(two lines)
The data output exports the measured value, date, and time.
Output Without stability* The data is output without the weight value being stable.
With stability The data is output only when the weight value is stable.
Automatic output Off* Deactivates the automatic data output. The data output must
be triggered by means of an SBI command with “ESC P”.
Each weight value Starts the automatic data output with each value according
tothe display result.
1 second Starts the automatic data output every second.
2 seconds Starts the automatic data output every 2seconds.
5seconds Starts the automatic data output every 5seconds.
10seconds Starts the automatic data output every 10seconds.
30seconds Starts the automatic data output every 30seconds.
60seconds Starts the automatic data output every 60seconds.
Display SBI error
messages
Off* Displays no error messages.
On Displays error messages if errors occur when receiving
andexecuting SBI commands.
Formatting output
values
Output in one line with
full length
Outputs the elements in the full length, e.g., header
andvalue.
The elements are separated by a space character, left-
aligned, and output without infilling with space characters.
Output in several lines
with 20 char. each
Distributes outputs, which due to their length do not fit in the
standard output format of 20 characters, over several lines
with 20 characters each.
Aligns each line to the left and infills with space characters
tomake up 20 characters.
Cut off to a
20characterline*
Defines the output in a standard output format
of20 characters.
Crops the end of longer outputs.
* Factory setting
7.7.15 Parameters in the “Connections / Climate Modules” Menu
Parameter Setting Values Explanation
Climate modules Display of connected
climate module
Activates climate modules if they are connected to the device.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA225S Series and is the answer not in the manual?

Sartorius Cubis MCA225S Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA225S Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals