Operating Instructions Cubis
®
MCA 83
System Settings
7.7.30 Parameters in the “Device Settings / Sound (Loudspeaker)” Menu
Parameter Setting Values Explanation
Touch sounds Activated Deactivates the acoustic signal for touch and keypad operation.
Deactivated* Activates the acoustic signal for touch and keypad operation.
Message sounds Activated Deactivates the acoustic signal for messages.
Deactivated* Activates the acoustic signal for messages.
Sound for
end-of-action
Activated Deactivates the acoustic signal for the completion of an action.
Deactivated* Activates the acoustic signal for the completion of an action.
* Factory setting
7.7.31 Parameters in the “Device Maintenance” Menu
Parameter Setting Values Explanation
Update firmware List of connectors and
then list of firmware
Starts a firmware update.
Not for conformity-assessed devices.
Install
QAPPCenter
List of connectors and
then list of QAPP
Centers
Updates the QAPP Center using the selected connector.
Install customized
QAPP
List of connectors and
then list of customized
QAPPs
Updates a customized QAPP using the selected connector.
Device
databackup
List of connectors Saves the device data on the selected connector.
Restore
devicedata
List of connectors Loads saved device data from the selected connector.
Export device data Role management Saves device role management for an administrator, user,
orguest on the selected connector.
User management Saves device user management for an administrator, user,
orgueston the selected connector.
Export overview Displays the list of connectors, e.g., a server or connected
USBstick.
Import device data List of connectors for
import
Loads the device data with role and user management,
e.g.,froma server or connected USB stick.
Reset to factory
settings
Restore factory settings Resets the device to factory settings and deletes
allconfigurations.
The audit trail and data memory will be retained.
Create
diagnosticfile
Saves information about device configuration and operation
on a connected USB stick.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA225S Series and is the answer not in the manual?

Sartorius Cubis MCA225S Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA225S Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals