Operating Instructions Cubis
®
 MCA  47
  Installation 
Requirements
 − The weighing pan and the associated components have not been set up.
 − For a device with an analytical draft shield or flat glass draft shield: 
Theside panels and upper panel have not been fitted.
Procedure
 t Turn the device on its side and place it on the soft support base.
 t Remove the hook for below-balance weighing (1) from the retainer on 
the underside of the base of the device.
 t Pull the cover of the below-balance weighing equipment (1) out.
 t NOTICE Damage to the device from cross-threading! Ensure that the 
hook for below-balance weighing is inserted straight into the thread (1).
 t Set up the device on the weighing table with recess.  
The hook for below-balance weighing may not touch the floor.
 t Install the draft protection shield.
1
1
1