Operating Menu Overview (Parameters)
o = Factory setting
W = User-defined setting
Setup
Application Parameters: Please refer to the “Basic Application Programs” manual
Fn key
Off
o Gross/net toggling
Toggle weight units
10 x higher resolution
Device Parameters
WP-1
Off
RS-232
1
)
SBI standard
SBI verifiable (trade)
o IS-232
ADC-232
Internal
Calibration/Adjustment
CAL Key Function
o Ext. cal./adjust.; default weight
Ext. cal./adjust.; weight can be selected
Key blocked
Calibration/Adjustment Sequence
Calibration with automatic adjustment
o Calibration with adjustment triggered manually
isoCAL Function
o Off
Adjustment prompt
External Calibration/Adjustment
2
)
o Accessible
Blocked
External Weight
Cal./adj. weight:
Adapt filter
Very stable conditions (“Minimal vibration”)
o Stable conditions (“Normal vibration”)
Unstable conditions (“Strong vibration”)
Very unstable conditions (“Extreme vibration”)
Application Filter
o Final readout
Filling mode
Low filtering
W/o filtering
Stability Range
1/4 digit
1/2 digit
o 1 digit
2
)
2 digits
2
)
4 digits
2
)
8 digits
2
)
1
) = function will be made available in future
2
) = not available on equipment verified for use in legal metrology